Getting Started
1- 12
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
To degauss and autobalance a current probe, perform these steps:
1.
Verify that the current probe is connected to the amplifier.
2.
Remove the current probe from the conductor under test.
3.
Lock the probe slide closed (see Figures 1--5 and 1--6).
4.
Press the amplifier
PROBE DEGAUSS AUTOBALANCE
button.
5.
Wait about five seconds for the degauss procedure to complete.
The PROBE DEGAUSS AUTOBALANCE LED glows green when the
operation has successfully completed. If the LED is blinking orange, the degauss
operation is still in progress. If the LED is red, the operation failed, and the
cause of the failure needs to be found and fixed. For more information, refer to
the PROBE DEGAUSS AUTOBALANCE button discussion on page 2--2.
NOTE
. The degauss procedure will fail if the amplifier is not properly connected
to an oscilloscope having 50
Ω
input impedance. If this occurs, the NOT
TERMINATED INTO 50
Ω
LED lights on the amplifier front panel.
After you have completed the oscilloscope adjustments and the amplifier
degauss/autobalance procedure, your system is ready to measure current.
DC Measurements
To measure DC current, perform these steps:
1.
Verify that the amplifier and the oscilloscope input coupling are set to DC,
and the input impedance is set to 50
Ω
.
2.
Lock the probe closed without a conductor passing through it.
3.
Adjust the ground reference of the oscilloscope to move the trace to the
desired graticule line.
4.
Press the amplifier
PROBE DEGAUSS AUTOBALANCE
button.
The NOT TERMINATED INTO 50
Ω
LED is lighted if impedance is not 50
Ω
.
If this is the case, make necessary changes. (For example, use a 50
Ω
termina-
tion.)
5.
After the degauss/autobalance routine completes, adjust the ground reference
(if necessary) using the amplifier
MANUAL BALANCE
controls.
Содержание TCPA300 SERIES
Страница 4: ......
Страница 8: ...Table of Contents iv TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 12: ...Table of Contents viii TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 18: ...Service Safety Summary xiv TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 21: ...Getting Started...
Страница 22: ......
Страница 37: ...Operating Basics...
Страница 38: ......
Страница 45: ...Reference...
Страница 46: ......
Страница 64: ...Application Notes 3 18 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 71: ...Specifications...
Страница 72: ......
Страница 86: ...Specifications 4 14 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 88: ......
Страница 89: ...Performance Verification...
Страница 90: ......
Страница 135: ...Adjustment Procedures...
Страница 136: ......
Страница 138: ...Adjustment Procedures Overview 6 2 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 146: ...TCP305 and TCP312 Adjustments 6 10 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 157: ...Maintenance...
Страница 158: ......
Страница 160: ...Customer Maintenance 7 2 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 181: ...Replaceable Parts...
Страница 182: ......
Страница 190: ...Amplifier Replaceable Parts 8 8 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 198: ...Current Probes Replaceable Parts 8 16 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 199: ...Glossary and Index...
Страница 200: ......