Adjustment Procedures
4- 2
RSA3408B Analyzer Service Manual
Equipment Required
Table 4--1 lists the test equipment required to adjust the RSA3408B analyzer.
Table 4- 1: Test equipment
Item number and
description
Minimum requirements
Example
Purpose
1. Signal generator
Frequency range: 10 kHz to 3.5 GHz;
Variable amplitude from --70 dBm to
+13 dBm into 50
Ω
;
Accuracy: <
±
1 dB; Function: FM
Agilent 8648D
Adjusting flatness in
baseband and RF1
2. Signal generator
Frequency range: 10 MHz to 8 GHz;
Output level: --30 dBm to +10 dBm;
Accuracy: <
±
1 dB
Agilent 83712B with
option 1E1
Adjusting flatness in
RF2 and RF3
3. Signal generator
Frequency range: 1 MHz to 8 GHz;
Output level: --30 dBm to +10 dBm;
Accuracy: <
±
1 dB
Agilent E8257D with
option 1E1
(option UNX recommended)
Adjusting flatness in all bands
4. Signal generator
Frequency: 6 GHz; Output:
≥
0dBm;
Digital modulation: QPSK
Agilent E4438C with
Options 506, UNJ, and 602
Adjusting wide/IQ flatness
5. Waveform generator
Arbitrary waveform; Function genera-
tor; Two output channels; Sampling
frequency: 50 kHz to 1 GHz; Ampli-
tude range: 20 mV to 2 V
Tektronix AWG520
1
Adjusting IQ gain
(Option 03 only)
6. Frequency counter
Frequency range: 10 MHz;
Aging rate: <5
×
10
--10
/day;
Accuracy: <0.01 ppm at 10 MHz
Agilent 53132A with
option 010
Checking the reference out-
put frequency accuracy
7. RF power meter
2
1 MHz to 8 GHz
Agilent 437B
3
Agilent E4418B
3
Adjusting the signal generator
output level
8. RF power sensor
2
10 MHz to 8 GHz; RF Flatness: <3%;
Uncertainty of calibration factor data:
<2% (RSS)
Agilent 8481A
Agilent E4412A
Adjusting the signal generator
output level
9. BNC cable
(five required)
50
Ω
, 36 in, male-to-male BNC
connectors
Tektronix part number
012-1341-XX
Signal interconnection
10. N-N cable
50
Ω
, 36 in, male-to-male N
connectors
Signal interconnection
11. GPIB cable
(three required)
2m, double-shielded
Tektronix part number
012-0991-XX
Software-based tests
12. Terminator
(two required)
Impedance: 50
Ω
; connectors: female
BNC input, male BNC output
Tektronix part number
011-0049-01
Signal termination for check-
ing frequency accuracy
13. RF attenuator
Ratio: 10 dB; impedance 50
Ω
Checking flatness
14. N adapter
2
Female-to-female N connectors
Signal interconnection
Содержание RSA3408B
Страница 4: ......
Страница 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Страница 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Страница 19: ...Specifications...
Страница 20: ......
Страница 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Страница 23: ...Theory of Operation...
Страница 24: ......
Страница 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Страница 31: ...Performance Verification...
Страница 32: ......
Страница 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Страница 89: ...Adjustment Procedures...
Страница 90: ......
Страница 127: ...Maintenance...
Страница 128: ......
Страница 173: ...Removal and Installation Procedures RSA3408B Analyzer Service Manual 5 45 Figure 5 23 Motherboard removal...
Страница 209: ...Diagrams...
Страница 210: ......
Страница 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Страница 214: ...6 4 RSA3408B Analyzer Service Manual...
Страница 216: ...6 6 RSA3408B Analyzer Service Manual...
Страница 217: ...Replaceable Parts List...
Страница 218: ......
Страница 229: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 11 1 2 RF4 RF2 RF3 RF6 3 4 5 RF5 6 Figure 7 4 RF modules...
Страница 237: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 19 1 2 3 4 5 6 7 8 9 10 11 Figure 7 8 Disk modules Option 06 2...
Страница 245: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 27 1 2 3 4 5 6 7 8 9 10 11 12 Figure 7 12 Chassis modules 2...
Страница 249: ......
Страница 250: ......