Performance Tests
RSA3408A Technical Reference
1- 27
Equipment
Required
One signal generator (Item 2)
One spectrum analyzer (Item 4)
One power meter (Item 6)
One power sensor (Item 7)
One 50
Ω
N-N coaxial cable (Item 10)
Prerequisites
Refer to page 1--10.
1.
Set up the power meter and sensor:
Do step 1 on page 1--19 to set up the power meter and sensor.
2.
Set the signal generator output:
a.
Hook up the instruments:
Connect the signal generator (Item 1) output through a 50
Ω
N-N coaxial
cable to the power sensor input. See the following figure.
Output
Signal generator
Power meter
Power sensor
b.
Set the signal generator controls:
Frequency
100 MHz
. . . . . . . . . . . . . . . .
Level
+10 dBm
. . . . . . . . . . . . . . . . . . . .
Check Level Linearity