Brief Procedures
1- 6
RSA3408A Technical Reference
Span
RBW
Setup display
Span setting
Figure 1- 2: Setup display
c.
Using the general purpose knob, change the span setting as listed in
Table 1--1 and check that the RBW is displayed correctly.
Table 1- 1: Span and RBW
Span
RBW
36 MHz
100 kHz
15 MHz
80 kHz
5 MHz
20 kHz
100 kHz
500 Hz
1 kHz
20 Hz
d.
Using the numeric keypad, set the span back to 36 MHz.
(Press
3
>
6
>
MHz
, in that order, on the keypad.)
7.
Check the reference level:
a.
Press the
AMPLITUDE
key on the front panel.
b.
Make sure that the reference level is set to 0 dBm with the
Ref Level
side key. Check that 0 dBm is displayed on the upper left side of the
graticule (see Figure 1--3).
c.
Use the general purpose knob to set the reference level to --30 dBm.
d.
Confirm that A/D OVERFLOW is indicated in the red box at the top
center of the screen. Make sure that --30 dBm is displayed on the upper
left side of the graticule and that the spectrum waveform is distorted as
shown in Figure 1--3.
e.
Using the numeric keypad, set the reference level back to 0 dBm.
(Press
0
>
ENTER
, in that order, on the keypad.)