Performance Tests
RSA3408A Technical Reference
1- 23
c.
Calibrate the analyzer gain:
H
Press
CAL
>
Calibrate Gain
.
H
Press
Amplitude
>
Auto Level
.
8.
Check the amplitude:
a.
Press the
PEAK
key on the front panel to place the marker on the peak.
b.
Check that the marker readout is within --20 dBm
0.5 dB.
9.
Disconnect the test equipment:
Disconnect the cable at the analyzer input.