Performance Veri
fi
cation
The following steps instruct you to assemble the test setup that includes the probe. (See Figure 32.) The system and probe
rise time (
t
s+p
) that you measure in step 24 is used to calculate the probe rise time (
t
p
) in step 25.
Figure 32: Test system rise time setup with probe
17.
Remove the TekConnect-to-SMA adapter from the test setup.
18.
Connect the probe to the 80A03 TekConnect probe interface.
19.
Connect the matched SMA cables to the probe SMA inputs and the 80E08 sampling head (Channels 7 and 8).
20.
Set the attenuation on the probe to 12.5X.
The test setup should now be connected as shown. (See Figure 32.)
21.
Expand the horizontal scale to help locate the step edge, and then adjust horizontal range to 50 ps/div while maintaining
the edge view. For a more stable measurement display, turn averaging on.
22.
Adjust the vertical scale to 100 mV/div, averaging on.
23.
Adjust the vertical and horizontal positioning to place the rising edge of the signal on the second vertical and center
horizontal graticule lines.
24.
Use the oscilloscope measurement capability to display rise time. Rise time is determined from the 10% and 90%
amplitude points on the waveform. Record the rise time as
t
s+p.
25.
Calculate the probe rise time using the following formula:
26.
Record the calculated probe rise time on the test record.
P7313SMA Technical Reference
45
Содержание P7313SMA
Страница 1: ...P7313SMA 13 GHz Differential Probe Technical Reference P071196800 071 1968 00...
Страница 2: ......
Страница 3: ...P7313SMA 13 GHz Differential Probe Technical Reference www tektronix com 071 1968 00...
Страница 6: ......
Страница 8: ...Table of Contents ii P7313SMA Technical Reference...
Страница 12: ...Preface vi P7313SMA Technical Reference...
Страница 34: ...Operating Basics 22 P7313SMA Technical Reference...