Termination Voltage Accuracy
These tests compare the termination control voltage that you apply (using the adjustment control for that termination voltage
mode), to the termination voltage output at the Vterm monitor jacks.
NOTE.
The Auto mode LED will
fl
ash when the probe inputs are open-circuit, or below a 50 mV threshold. If the LED
continues to
fl
ash after you connect the inputs, cycle the mode SELECT button.
External (Ext) Mode
The Ext mode test setup is shown. (See Figure 24.)
1.
Plug the probe directly into an oscilloscope channel and set the Vterm Source Select to EXT on the probe.
Figure 24: Termination Voltage Accuracy, Ext mode setup
2.
Connect the 50
Ω
terminations on the three probe SMA connectors. This sets the common mode input voltage to 0.0 V.
3.
The probe attenuation can be set to either 2.5X or 12.5X.
4.
Using the 0.080 in pin-to-Banana plug cables, connect the power supply to the external DC input jacks on the front
of the probe.
5.
Set the power supply as close as practical to 0.000 volts, and use the DMM to measure this input voltage at the terminals
on the front of the probe. Record this voltage as Vin on the test record.
36
P7313SMA Technical Reference
Содержание P7313SMA
Страница 1: ...P7313SMA 13 GHz Differential Probe Technical Reference P071196800 071 1968 00...
Страница 2: ......
Страница 3: ...P7313SMA 13 GHz Differential Probe Technical Reference www tektronix com 071 1968 00...
Страница 6: ......
Страница 8: ...Table of Contents ii P7313SMA Technical Reference...
Страница 12: ...Preface vi P7313SMA Technical Reference...
Страница 34: ...Operating Basics 22 P7313SMA Technical Reference...