Performance Veri
fi
cation
Input Resistance
This test checks the differential mode input resistance (the resistance between each SMA input). The test is performed with
the probe disconnected from the oscilloscope.
1.
Disconnect the probe from the oscilloscope.
2.
Zero the DMM with its measurement leads connected together on the lowest scale that can measure 100
Ω
.
3.
Remove the SMA terminations from the two probe inputs and gently probe the center contacts of the input connectors.
Be careful not to touch the outer edge (ground) of the connector. (See Figure 23.)
4.
Measure the resistance and write down the value.
5.
Reverse the DMM connections and repeat the measurement. Write down the value.
6.
Add the two measurements from steps 4 and 5, and divide the total by two. Record the result in the test record.
7.
Connect the probe to the oscilloscope channel that you will use in the next test so that the probe warms up to operating
temperature.
Figure 23: Checking differential mode input resistance
P7313SMA Technical Reference