Performance Veri
fi
cation
Auto Mode
In Auto mode, the probe measures the input signal DC common mode voltage and automatically sets the termination voltage
to equal that voltage. In this test, the two signal inputs are connected together and driven by an external power supply to
set the common mode voltage to the 0.0, +2.500 and -2.500 volt test values.
1.
Connect the test setup as shown. (See Figure 25.)
Figure 25: Termination Voltage Accuracy, Auto mode setup
2.
Set the Vterm Source Select to Auto on the probe.
3.
Set the power supply as close as practical to 0.000 volts, and use the DMM to measure this input voltage at the terminals
on the power supply. Record this voltage as Vin on the test record.
4.
Use the second DMM to measure the output voltage at the termination voltage monitor jacks on the top of the probe.
Record this voltage as Vout on the test record, and verify that the Vout voltage is within the speci
fi
ed limits in the
min/max columns.
5.
Repeat steps 3 and 4 for the +2.500 volt and -2.500 volt input values listed in the test record.
38
P7313SMA Technical Reference