Appendix A: Alternate Verification Procedures
40
P6248 1.7 GHz Differential Probe Service Manual
Preparation
Allow all test equipment to warm up for 20 minutes in an environment that is
within the environmental conditions specified in Table 1 on page 1. Prepare the
equipment as follows:
1.
Connect the probe tip adapter to the inputs of the probe. (For a description of
these accessories, refer to the list of standard accessories on pages 33
and 34).
2.
Attach the probe to the 1103 TEKPROBE power supply.
3.
If necessary, set the line selector of the 1103 power supply to the correct
voltage.
4.
Turn on the 1103 power supply.
5.
Set the 1103 offset to off (button not illuminated).
6.
Allow at least 20 minutes for the probe and equipment to warm up.
Perform the verification procedures in order.