Performance Verification
P6248 1.7 GHz Differential Probe Service Manual
17
8.
If necessary, set the input impedance to 1 M
Ω
(bottom button along right
side of bezel).
9.
Watch the input impedance setting while attaching the probe to the CH 1.
10.
Verify that the input impedance changed to 50
Ω
.
11.
Set the probe attenuation to
÷
10.
12.
Verify that the channel 1 scale factor is 1 V/div.
13.
Set the probe attenuation to
÷
1.
14.
Verify that the channel 1 scale factor is 100 mV/div.
This completes the functional check and the performance verification.