Adjustment Procedures
P6248 1.7 GHz Differential Probe Service Manual
23
5.
Adjust the power supply to output about 700 mV.
6.
Adjust the
÷
1 gain adjustment until the output voltage measures the same as
the input within
±
5 mV.
7.
Change the probe attenuation to
÷
10.
8.
Change the DMM monitoring the input to the 20 or 30 volt range.
9.
Adjust the power supply to output approximately 7 V.
10.
Adjust the
÷
10 gain adjustment until the DMM measuring the output reads
1/10 of the same voltage as the DMM monitoring the input within
±
5 mV.
11.
Keep the probe output connections for the next step.
Offset (Final)
NOTE
. The offset and gain adjustments interact.
1.
Repeat steps 1 through 6 of the
Offset (preliminary)
procedure.
2.
Remove the output cable termination and input for the next procedure.
DC CMRR
1.
Set up the equipment as shown in Figure 12:
a.
Connect a BNC cable from the output of the 1103 to the Channel 1 input
of the oscilloscope. If the oscilloscope does not have 50
Ω
input setting,
add a 50
Ω
inline BNC terminator at the oscilloscope input.
b.
Connect the Y lead adapter to the probe input. Using square pins,
connect both leads to the red binding post of a binding post to BNC
adapter. Connect the BNC adapter to the function generator output. (If
necessary, use a BNC cable and a female-to-female BNC connector
between the BNC adapter and the generator.) If the function generator
reference (shield of the BNC connector) is isolated from ground, connect
the ground lead from the probe case to the black binding post.
NOTE
. Do not use the BNC to probe tip adapter for this connection. The power
level of the generator will exceed the terminator rating.
Содержание P6248
Страница 4: ......
Страница 10: ...Service Safety Summary vi P6248 1 7 GHz Differential Probe Service Manual...
Страница 12: ...Preface viii P6248 1 7 GHz Differential Probe Service Manual...
Страница 38: ...Adjustment Procedures 26 P6248 1 7 GHz Differential Probe Service Manual...
Страница 42: ...Options 30 P6248 1 7 GHz Differential Probe Service Manual...
Страница 43: ...P6248 1 7 GHz Differential Probe Service Manual 31 Replaceable Parts 1 2 3 4 Figure 15 P6248 replaceable parts...
Страница 50: ...Replaceable Parts 38 P6248 1 7 GHz Differential Probe Service Manual...
Страница 56: ...Appendix A Alternate Verification Procedures 44 P6248 1 7 GHz Differential Probe Service Manual...
Страница 57: ......