P6248 1.7 GHz Differential Probe Service Manual
1
Specifications
The specifications in Tables 1 through 4 apply to a P6248 differential probe
installed on a Tektronix TDS 694C oscilloscope. When the probe is used with
another oscilloscope, the oscilloscope must have an input impedance of 50
Ω
.
The probe must have a warm-up period of at least 20 minutes and be in an
environment that does not exceed the limits described in Table 1. Specifications
for the P6248 differential probe fall into three categories: warranted, typical, and
nominal characteristics.
Warranted Characteristics
Warranted characteristics (Table 1) describe guaranteed performance within
tolerance limits or certain type-tested requirements. Warranted characteristics
that have checks in the
Performance Verification
section are marked with the
n
symbol.
Table 1: Warranted electrical characteristics
Characteristic
Description
n
DC attenuation accuracy
±
2% for input voltages
±
750 mV (
÷
1) and
±
7.5 V (
÷
10)
±
5% for input voltages
±
850 mV (
÷
1) and
±
8.5 V (
÷
10)
Maximum non destructive input voltage
±
25 V
(DC + peak AC)
between signal and common of the same channel
n
Differential signal range (DC coupled)
±
8.5 V at attenuation setting of
÷
10
±
850 mV at attenuation setting of
÷
1
Common-mode signal range (DC coupled)
±
7 V
n
Output offset voltage
±
10 mV (+20 to +30
_
C, +68 to +86
_
F)
±
50 mV (0 to +50
_
C, +32 to +122
_
F)
n
Rise time (probe only)
<265 ps
n
Common-mode rejection ratio
45 dB at 1 MHz in
÷
10
j
25 dB at 100 MHz in
÷
10
20 dB at 500 MHz in
÷
10
20 dB at 500 MHz in
÷
10
18 dB at 1 GHz in
÷
10
18 dB at 1 GHz in
÷
10
60 dB at 1 MHz in
÷
1
38 dB at 100 MHz in
÷
1
38 dB at 100 MHz in
÷
1
35 dB at 500 MHz in
÷
1
35 dB at 500 MHz in
÷
1
30 dB at 1 GHz in
÷
1
Bandwidth (probe only)
DC to
≥
1.5 GHz (-- 3dB)
Temperature
Operating: 0 to +50
_
C (+32 to +86
_
F),
0 to +60
_
C (+32 to +122
_
F)
1
(probe head)
Nonoperating: --55 to +75
_
C (--131 to +167
_
F)
1
Содержание P6248
Страница 4: ......
Страница 10: ...Service Safety Summary vi P6248 1 7 GHz Differential Probe Service Manual...
Страница 12: ...Preface viii P6248 1 7 GHz Differential Probe Service Manual...
Страница 38: ...Adjustment Procedures 26 P6248 1 7 GHz Differential Probe Service Manual...
Страница 42: ...Options 30 P6248 1 7 GHz Differential Probe Service Manual...
Страница 43: ...P6248 1 7 GHz Differential Probe Service Manual 31 Replaceable Parts 1 2 3 4 Figure 15 P6248 replaceable parts...
Страница 50: ...Replaceable Parts 38 P6248 1 7 GHz Differential Probe Service Manual...
Страница 56: ...Appendix A Alternate Verification Procedures 44 P6248 1 7 GHz Differential Probe Service Manual...
Страница 57: ......
Страница 58: ......