Performance Verification
10
P6248 1.7 GHz Differential Probe Service Manual
7.
Verify that the attenuation is in the range of 9.5 to 10.5.
8.
Remove the Y-lead adapter from the probe and reverse the connection (black
to +, red to --).
9.
Measure and record the output voltage as V
out
.
10.
Calculate attenuation as V
in
÷
(-- V
out
).
11.
Verify that the attenuation is in the range of 9.5 to 10.5.
12.
Set the differential mode source to 0.85 V
±
10 mV. Record the actual voltage
as V
in
.
13.
Make sure that the common mode source is still set to 7.0 V
±
50 mV.
14.
Set the probe attenuation to
÷
1.
15.
Measure and record the output voltage as V
out
.
16.
Calculate attenuation as V
in
÷
(-- V
out
).
17.
Verify that the attenuation is in the range of 0.95 to 1.05.
18.
Remove the Y-lead adapter from the probe and reverse the connection again
(red to +, black to --).
19.
Measure and record the output voltage as V
out
.
20.
Calculate the attenuation as V
in
÷
V
out
.
21.
Verify that the attenuation is in the range of 0.95 to 1.05.
22.
Remove all connections.
Содержание P6248
Страница 4: ......
Страница 10: ...Service Safety Summary vi P6248 1 7 GHz Differential Probe Service Manual...
Страница 12: ...Preface viii P6248 1 7 GHz Differential Probe Service Manual...
Страница 38: ...Adjustment Procedures 26 P6248 1 7 GHz Differential Probe Service Manual...
Страница 42: ...Options 30 P6248 1 7 GHz Differential Probe Service Manual...
Страница 43: ...P6248 1 7 GHz Differential Probe Service Manual 31 Replaceable Parts 1 2 3 4 Figure 15 P6248 replaceable parts...
Страница 50: ...Replaceable Parts 38 P6248 1 7 GHz Differential Probe Service Manual...
Страница 56: ...Appendix A Alternate Verification Procedures 44 P6248 1 7 GHz Differential Probe Service Manual...
Страница 57: ......
Страница 58: ......