Performance Verification
P6205 Instruction Manual
4–9
Aberrations Check
Use the following procedure to confirm that the P6205 probe
performs within its Aberrations specifications limits.
1. Set the oscilloscope controls as indicated in the table below.
Oscilloscope settings
Control
Setting
Volts/division
50 mV
Time/division
500 ps
Trigger coupling
DC
Trigger source
Internal clock
Trigger mode
Auto
Waveform averaging
On
Number of waveforms to average
8
2. Install one SMA-to-BNC adapter on an SD-24 sampling head
input channel and another adapter on the oscilloscope calibrator
output. Use the 10-inch coaxial cable to connect the oscilloscope
calibrator output to the SD-24 input channel. See Figure 4–7
below.
11801B oscilloscope
Figure 4–7: Aberrations test setup part 1
Содержание P6205
Страница 4: ......
Страница 12: ...Service Safety Summary viii P6205 Instruction Manual...
Страница 15: ...Getting Started...
Страница 16: ......
Страница 24: ...Getting Started 1 8 P6205 Instruction Manual...
Страница 25: ...Operating Basics...
Страница 26: ......
Страница 32: ...Operating Basics 2 6 P6205 Instruction Manual...
Страница 33: ...Specifications...
Страница 34: ......
Страница 41: ...Performance Verification...
Страница 42: ......
Страница 57: ...Maintenance...
Страница 58: ......
Страница 65: ...Replaceable Parts...
Страница 66: ......
Страница 72: ...Replaceable Parts 6 6 P6205 Instruction Manual Figure 6 2 P6205 probe optional accessories...
Страница 74: ...Replaceable Parts 6 8 P6205 Instruction Manual...