Performance Verification
4–6
P6205 Instruction Manual
1. Set the oscilloscope controls as indicated in the table below.
Oscilloscope settings
Control
Setting
Volts/division
5 mV
Time/division
500 ps
Trigger coupling
DC
Trigger source
Internal clock
Trigger mode
Auto
Waveform averaging
On
Number of waveforms to average
8
2. Connect one end of the 10-inch coaxial cable to the channel 1
output of the 1103 probe power supply. Use a male SMA-to-BNC
female adapter to connect the free end of the cable to an SD-24
sampling head input. See Figure 4–5 below.
1103 power supply
11801B
oscilloscope
Figure 4–5: Rise time test setup
3. Press SELECT CHANNEL on the sampling head to enable the
corresponding input channel.
4. Connect the P6205 probe to the 1103 probe power supply
channel 1 input. The P6205 does not use the offset capability of
Содержание P6205