Performance Verification
4–2
P6205 Instruction Manual
Table 4–1: Test equipment
Description
Minimum Requirement
Example Product
Oscilloscope
DC to 10 GHz, 50
W
input
Tektronix 11801B with
SD-24 TDR sampling head
Probe power supply
TEKPROBE interface con-
nector
Tektronix 1103 TEKPROBE
power supply
Calibration generator
Amplitude accuracy:
±
0.25% Calibration signal from
11801B oscilloscope
Pulse generator
Rise time: <100 ps
Calibration signal from
11801B oscilloscope
Termination adapter
50
W
, probe-to-GR
Tektronix part number
017-0088-00
Female BNC-to-GR adapter
Tektronix part number
017-0063-00
Male SMA-to-BNC female
adapters (2)
Tektronix part number
015-0554-00
Male SMA-to-GR adapter
Tektronix part number
015-1007-00
Coaxial cable
50
W
BNC, 10 inch length
Tektronix part number
012-0208-00
Attenuation Check
Use the following procedure to confirm that the P6205 probe
performs within its Attenuation specification limits.
1. Set the oscilloscope controls as indicated in the table below.
Oscilloscope settings
Control
Setting
Volts/division
50 mV
Time/division
2
m
s
Содержание P6205
Страница 4: ......
Страница 12: ...Service Safety Summary viii P6205 Instruction Manual...
Страница 15: ...Getting Started...
Страница 16: ......
Страница 24: ...Getting Started 1 8 P6205 Instruction Manual...
Страница 25: ...Operating Basics...
Страница 26: ......
Страница 32: ...Operating Basics 2 6 P6205 Instruction Manual...
Страница 33: ...Specifications...
Страница 34: ......
Страница 41: ...Performance Verification...
Страница 42: ......
Страница 57: ...Maintenance...
Страница 58: ......
Страница 65: ...Replaceable Parts...
Страница 66: ......
Страница 72: ...Replaceable Parts 6 6 P6205 Instruction Manual Figure 6 2 P6205 probe optional accessories...
Страница 74: ...Replaceable Parts 6 8 P6205 Instruction Manual...