Adjustments
P6015A Instruction Manual
2–7
R9
DC
ATTEN
MID 3
C4
C2
C5
COMP
R5
R2
R4
C1
MID 1
MID 2
LF
R6
HF
COMP
R7
C7
Figure 2–2: Adjustment Locations
Low-Frequency Compensation
1. Set the calibration generator to high amplitude output, and set the
period to 1 ms. Set the pulse amplitude to display five divisions.
2. Set the oscilloscope time base to display 200
m
s/div.
3. Center the waveform on the screen.
4. The leading corner of the square wave should be level with the
trailing corner. If necessary, adjust
C5
.
5. The top of the waveform should be flat to within
±
5% (
±
1.25 mi-
nor divisions). If necessary, perform steps 6 through 8.