In Detail
P6015A Instruction Manual
1–13
Figure 1–4: Maximum Input Voltage Derating (DC + Peak AC)
20 kV
28 kV
33 kV
36 kV
40 kV
10% Duty Cycle 100 ms on time
1 sec
100 ms
10 ms
1 ms
Duration
Maximum V
oltage
(peak)
20% Duty Cycle 50 ms on time
30% Duty Cycle 40 ms on time
100% Duty Cycle 100 ms on time
50% Duty Cycle 30 ms on time
Figure 1–5: Peak Pulse Derating
Probe Grounding
When making any measurement, use the probe ground connector to
form the basic two-terminal connection to the device under test.
CAUTION.
The ground lead is rated at 1 kV maximum. Make sure the
ground lead does not contact the probe tip or a high-voltage point on
the circuit under test.