Specifications
P6015A Instruction Manual
1–25
Table 1–4: Warranted Environmental Characteristics
Characteristic
Information
Temperature
Nonoperating
–55
_
C to +75
_
C (–67
_
F to +167
_
F)
Operating
DC + peak AC
Peak Pulse
0
_
C to +35
_
C (+32
_
F to +95
_
F)
0
_
C to +50
_
C (+32
_
F to +122
_
F)
(See Table 1–1 on page 1–10 and Time Limitations
Specification below)
Humidity
Nonoperating / Operating
95% relative humidity at +50
°
C (+122
°
F). See
Figure 1–7 for derating characteristics.
Maximum altitude
Nonoperating
15,000 m (50,000 ft)
Operating
4,600 m (15,000 ft)
Peak pulse voltage derated from 40 kV at 8000 feet
(2440 m) to 30 kV at 15,000 feet (4570 m).
Vibration (random)
Nonoperating
3.48 g rms from 5 to 500 Hz. Ten minutes on each axis.
Operating
2.66 g rms from 5 to 500 Hz. Ten minutes on each axis.
Shock (nonoperating)
500 g, half sine, 0.5 ms duration, 18 shocks total in three
axis.
Time Limitations
Less than 70% of Rated
Input Voltage at 0–35
_
C
No time limit
Greater than 70% of
Rated Input Voltage at
0–35
_
C
30 minutes maximum
in any 2.5 hour period
35–50
_
C
15 minutes maximum
in any 2.5 hour period
Содержание P6015A
Страница 4: ......
Страница 12: ...Service Safety Summary viii P6015A Instruction Manual...
Страница 14: ...Service Safety Summary x P6015A Instruction Manual...
Страница 15: ...User Information...
Страница 16: ......
Страница 18: ...Overview 1 2 P6015A Instruction Manual Guard Ring Figure 1 1 The P6015A High Voltage Probe...
Страница 38: ...In Detail 1 22 P6015A Instruction Manual...
Страница 46: ......
Страница 47: ...Service Information...
Страница 48: ......
Страница 50: ...Performance Verification 2 2 P6015A Instruction Manual...
Страница 68: ...Replaceable Parts 2 20 P6015A Instruction Manual 1 2 3 4 5 6 7 8 9 10 Figure 2 5 P6015A Exploded View...
Страница 73: ...Index...
Страница 74: ......