Performance Verification
32
OI1125 E/O Transmitter Instruction Manual
Prerequisites
The tests in this section comprise an extensive, valid confirmation of perfor-
mance and functionality when the following requirements are met:
H
The instrument must have been last calibrated and adjusted at an ambient
temperature b20
_
C and +30
_
C, must have been operating for a
warm-up period of at least 30 minutes, and must be operating in an
environment with temperature, altitude, humidity, and vibration within the
operating limits described in Environmental characteristics on page 21.
H
The test equipment (oscilloscope, pattern generator, tunable laser and optical
meters), must be operating for a warm-up period of at least 30 minutes.
Equipment Required
Table 11 lists the equipment required to perform the performance verification
procedure. The types and quantities of connectors may vary depending on the
specific equipment you use.
CAUTION.
To prevent damage to your equipment and cables, use the proper type
optical connectors. The optical connection styles shown in the test setups are
correct for the equipment called out in Table 11 below. If you use alternative
equipment, verify the connection style of the alternative equipment, and use the
appropriate cable or adapter.
Table 11: Test equipment
Description
Minimum requirements
Example product
Sampling oscilloscope
with optical sampling
head
Bandwidth: 20 GHz with limiting filters
for OC 48, OC192, and 12.5 Gb/s
Jitter:
t
1 ps at min samp delay
Tektronix TDS8000 with
80C01 optical sampling
head
Pattern generator
t
2.4 Gb/s to
u
12.5 Gb/s,
PRBS 2
31–1
Advantest D3186 with
options 13 and 72
Wavelength meter
1530 nm to 1565 nm
Burleigh WA-1100
Optical power meter with
optical power head
1550 nm, +15 dBm to –40 dBm,
0.01 dBm resolution,
±
5% accuracy
Advantest Q8221 with
Q82227 power head
Tunable Laser
10 dBm output power over 1530 nm to
1565 nm range,
PMF-style output connector
Tunics BT-1560 w/high
power option
Cable, low-loss coax (2)
50
W
, SMA male each end, ~1 m
Tektronix part number
015-0561-00
Cable, coax
50
W
, SMA male each end, 1 m
Tektronix part number
174-1341-00
Содержание OI1125
Страница 4: ......
Страница 8: ...Table of Contents iv OI1125 E O Transmitter Instruction Manual...
Страница 20: ...Getting Started 8 OI1125 E O Transmitter Instruction Manual...
Страница 30: ...Operating Basics 18 OI1125 E O Transmitter Instruction Manual...
Страница 36: ...Specifications 24 OI1125 E O Transmitter Instruction Manual...
Страница 38: ......
Страница 42: ...Theory of Operation 30 OI1125 E O Transmitter Instruction Manual...
Страница 72: ...Performance Verification 60 OI1125 E O Transmitter Instruction Manual...