Performance Verification
OI1125 E/O Transmitter Instruction Manual
45
Oscilloscope Setup.
Set the TDS8000 Vertical settings as follows:
NOTE
. Due to differences in oscilloscope firmware, your setup steps may differ
slightly.
1. From the toolbar, select Setup
→
Vertical.
2. Click the Optical>> button.
3. Set the Signal Conditioning mode:
a. In the Wavelength field, select 1550 from the drop-down menu.
b. In the Filter field, select OC-48 from the drop-down menu.
NOTE
. You must change the Filter field setting when you take measurements at
other data rates. See Table 13.
A filter is not used for the 12.50 Gb/s data rate. Instead, set the Filter field to
None, and set the bandwidth to 12.50 GHz.
Table 13: Signal conditioning mode settings
Data rate,
Gb/s
Wavelength
field setting
Filter field
setting
Bandwidth
setting, GHz
2.48832
1550
OC-48
–––––––
9.95328
1550
OC-192
–––––––
12.50000
1550
None
12.50
4. Select the Trig tab to set the trigger parameters.
a. In the Trigger Source box, select External Direct.
Preview the Eye Pattern.
1. Turn the BERT data and clock outputs on.
2. Enable the transmitter.
3. Turn on the channel that the optical module is plugged into.
4. Press Autoset and then adjust the horizontal scaling to display an eye pattern.
The oscilloscope displays an eye pattern as shown in Figure 17 on page 47.
Содержание OI1125
Страница 4: ......
Страница 8: ...Table of Contents iv OI1125 E O Transmitter Instruction Manual...
Страница 20: ...Getting Started 8 OI1125 E O Transmitter Instruction Manual...
Страница 30: ...Operating Basics 18 OI1125 E O Transmitter Instruction Manual...
Страница 36: ...Specifications 24 OI1125 E O Transmitter Instruction Manual...
Страница 38: ......
Страница 42: ...Theory of Operation 30 OI1125 E O Transmitter Instruction Manual...
Страница 72: ...Performance Verification 60 OI1125 E O Transmitter Instruction Manual...