Performance Verification
OI1125 E/O Transmitter Instruction Manual
49
8. Set the Filter field on the oscilloscope to None.
9. Set the bandwidth field on the oscilloscope to 12.50 GHz. See steps 1
through 3 in the Oscilloscope Setup on page 45.
10. Repeat steps 1 through 3 for the 12.50 Gb/s, 1.5 V row in Table 18.
11. Disconnect the optical cable from the oscilloscope and perform another dark
level compensation. Reconnect the cable when the compensation is done.
12. Disconnect the cable from the BERT output and connect the 2X attenuator to
the BERT output. Connect the cable to the attenuator.
13. Set the BERT amplitude to 0.5 V.
14. Repeat steps 1 through 3 for the 12.50 Gb/s, 0.25 V row in Table 18.
15. Complete the remaining 0.25 V rows in Table 18, making sure to set the
proper filter field on the oscilloscope for each row.
16. Keep the test setup connected for the next test.
Содержание OI1125
Страница 4: ......
Страница 8: ...Table of Contents iv OI1125 E O Transmitter Instruction Manual...
Страница 20: ...Getting Started 8 OI1125 E O Transmitter Instruction Manual...
Страница 30: ...Operating Basics 18 OI1125 E O Transmitter Instruction Manual...
Страница 36: ...Specifications 24 OI1125 E O Transmitter Instruction Manual...
Страница 38: ......
Страница 42: ...Theory of Operation 30 OI1125 E O Transmitter Instruction Manual...
Страница 72: ...Performance Verification 60 OI1125 E O Transmitter Instruction Manual...