Performance Veri
fi
cation
3.
Open the
test40.TRP
fi
le on the MTX100B.
a.
Select
File
>
Open
in the Play screen to open the
Select File
dialog box.
b.
In the dialog box, select the
test40.TRP
fi
le.
4.
Press the
Play/Pause
button on the MTX100B to start playing the test40.TRP
fi
le.
5.
Use the oscilloscope to measure that the high and low levels of the Event
output are as follows:
Control
Setting
High level
> 2.4 V
Low level
< 0.4 V
6.
Press the
Stop
button on the MTX100B to stop the stream output.
IEEE1394/ASI Interface
(Option 05 Only)
This test veri
fi
es that transport stream data is correctly output from and is recorded
through the IEEE1394/ASI interface on the MTX100B.
Equipment required.
MPEG analyzer
Oscilloscope
D-VHS digital recorder
Video monitor
75
Ω
BNC cable
IEEE1394 cable (4 pin-4 pin)
75
Ω
signal adapter
RCA(Ma)-to-BNC(Fe) adapter
test40.TRP
fi
le
2–36
MTX100B Speci
fi
cations and Performance Veri
fi
cation