Performance Veri
fi
cation
MTX100B Test Record
Print this page and use it to record the performance test result.
Table 2-2: MTX100B test record
Serial Number:
Cal Date:
Temperature:
Humidity:
Performance Test
Minimum
Measured
Maximum
Internal Clock Frequency
26.999924 MHZ
27.000076 MHz
1
Test record limits are based on the SMPTE 310M frequency accuracy requirement.
Procedures
The following conventions are used in the performance veri
fi
cation procedures:
Each test lists the characteristic that is being tested and the equipment required
to perform the test.
Each test contains complete setup instructions, which allows you to perform
each test individually or in order.
The equipment connection illustrations are speci
fi
c to the recommended
equipment (equipment nomenclature is labeled). If you are using substitute
equipment, the location of your signal connections may vary from those in
the illustrations.
NOTE.
Before you begin the performance veri
fi
cation procedures, be sure that
the MTX100B is operating in an environment that is within the operating limits.
(See Table 1-13 on page 1-17.)
In addition, the MTX100B and the test equipment must be warmed up for at least
20 minutes to ensure accurate test results.
MTX100B Speci
fi
cations and Performance Veri
fi
cation
2–7
Содержание MTX100B
Страница 2: ......
Страница 6: ......
Страница 12: ...General Safety Summary vi MTX100B Specifications and Performance Verification ...
Страница 16: ...Introduction x MTX100B Specifications and Performance Verification ...
Страница 17: ...Specifications ...
Страница 18: ......
Страница 36: ...Specifications 1 18 MTX100B Specifications and Performance Verification ...
Страница 37: ...Performance Verification ...
Страница 38: ......