Performance Veri
fi
cation
3.
Open the
test40.TRP
fi
le on the MTX100B.
a.
Select
File
>
Open
to open the
Select File
dialog box.
b.
In the dialog box, select the
test40.TRP
fi
le.
4.
Press the
Play/Pause
button on the MTX100B to start playing the
test40.TRP
fi
le.
5.
Use the oscilloscope to verify that the amplitude and rise and fall times are
within the following range.
Control
Setting
Amplitude
720 mV to 880 mV
Rise and fall time
≤
1.2 ns
6.
Disconnect the 75
Ω
BNC cable from the MTX100B and the oscilloscope.
2–38
MTX100B Speci
fi
cations and Performance Veri
fi
cation