The next figure shows how a floor mounting bracket is installed. Keithley part numbers are included for the required
hardware.
Figure 29: S500 system cabinet floor mount
Optional advanced seismic securement
If you ordered a system with the advanced seismic fastening, follow these mounting instructions.
1. Position the cabinet in the desired location on a smooth, level floor.
2. Place the floor mounting brackets at the corners and make sure you have enough room for proper placement.
3. To properly place the mounting brackets, the cabinet must be lifted by the leveling legs.
Lifting the cabinet will allow the holes in the mounting brackets to line up with the holes in the plinth.
4. Temporarily attach the mounting brackets to each corner with the provided screws (see next figure).
The next figure shows the restraint brackets and bolt installation for the system cabinet.
Figure 30: S500 seismic restraints