Section 3: Setting up SMUs in Clarius
Model 4200A-SCS Source-Measure Unit (SMU) User's Manual
3-34
4200A-SMU-900-01 Rev. A December 2020
Figure 52: Select Output Values
Compliance exit-condition options
The compliance exit-condition options are available from the Test Settings pane when you select
Exit
Condition
.
When the source goes into compliance, you can choose the action that the system takes.
You can select one of the following exit-condition options:
•
None
: The run continues. This is the default setting.
•
Test
: The 4200A-SCS exits the test that is presently being run. If there are additional tests,
operation continues to the next test.
•
Device
: The 4200A-SCS exits the device that is presently being run. If there are additional
devices, operation continues to the next device.
•
Subsite
: The 4200A-SCS exits the subsite presently being run. If there are additional subsites,
operation continues to the next subsite.
•
Site
: The 4200A-SCS exits the site. If there are multiple sites, operation continues to the next
site.
•
Project
: The 4200A-SCS exits the project. All testing stops.
To set the Exit Condition when compliance occurs:
1. Select the test in the project tree.
2. Select
Configure
.
3. Select
Test Settings
.
4. Select
Exit Condition
.
5. Select the option for your test.
6. Select
OK
.