PA CNTRL connector
The PA CNTRL (preamplifier control) terminal is a 15-pin D-sub connector that provides both power
and signal connections to the 4200-PA Remote Preamplifier. The preamplifier can either be mounted
and connected directly to the SMU, or it can be connected to the SMU using a cable (4200-RPC-
X
)
when mounted remotely. Refer to the
Source-measure unit (SMU) with 4200-PA overview
4-5) for more information on the preamplifier.
Basic SMU and preamplifier circuit configuration
A basic SMU and preamplifier circuit configuration is shown in the following figure.
The preamplifier FORCE terminal is connected to DUT HI. DUT LO is connected to GROUND.
Figure 24: Basic SMU and preamplifier source-measure configuration