Performance tests
NOTE.
This test uses the .wfmx
fi
les that are used in the previous test (Analog
Amplitude Accuracy). If you did not perform that test, use step 7 in the procedure
to load the
fi
les, and then proceed with the following steps to activate the
8+2 markers.
5.
In the
Waveform List
window, select (drag and drop) the
PV_DC_Plus.wfmx
waveform to the work space.
6.
Press the
Home
button on the instrument, or click the
Home
tab on the display.
a.
Click on the
Setup
tab.
b.
From the drop-down list under Resolution (bits), select
8+2 Mkrs
.
7.
Click the CH 1
Enable
button on the instrument to enable the channel 1
output.
8.
Press the
All Outputs
button on the instrument to output the waveform.
9.
Make the instrument High Level setting shown in the
fi
rst row of the
following table:
Table 2-7: Marker high level accuracy
High level settings
Accuracy limits
+ 1.4 V
1.185 V to 1.615 V
0.0 V
–75 mV to +75 mV
–0.9 V
–1.065 V to –0.735 V
10.
Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_1
.
11.
Use the following formula to compensate the voltage for the 50
Ω
BNC
termination:
Marker_High = ( 50) / (2 Term_R) Measured_voltage_1
Where Term_R is the resistance of the 50
Ω
BNC termination measured in
step 3 in the
Measure the termination resistance
procedure. (See page 2-24,
Measure the termination resistance
12.
Verify that the marker High level falls within the limits given in the table.
(See Table 2-7.)
13.
Repeat steps 9 through 12 for the remaining 2 rows in the table. (See
Table 2-7.)
2–32
AWG70000A Series and AWGSYNC01 Technical Reference
Содержание AWG70001A
Страница 2: ......
Страница 12: ...Service safety summary viii AWG70000A Series and AWGSYNC01 Technical Reference...
Страница 14: ...Preface x AWG70000A Series and AWGSYNC01 Technical Reference...
Страница 15: ...Specifications...
Страница 16: ......
Страница 42: ...AWGSYNC01 1 26 AWG70000A Series and AWGSYNC01 Technical Reference...
Страница 43: ...Performance verification...
Страница 44: ......