Performance tests
Marker high and low level accuracy
Required equipment
Prerequisites
Digital multimeter
BNC-dual banana adapter
50
Ω
BNC cable
50
Ω
BNC termination
SMA male-BNC female adapter
50
Ω
SMA termination
+ Marker high level
accuracy
1.
Perform the
Measure the termination resistance
procedure. (See page 2-24,
Measure the termination resistance
2.
Press the
All Outputs On/Off
button on the instrument to turn off all the
outputs.
3.
Use the 50
Ω
SMA cable, SMA-BNC adapter, 50
Ω
BNC termination, and
BNC-Banana adapter to connect the CH 1 Markers M1 + connector on the
instrument to the HI and LO inputs on the digital multimeter.
4.
Use the 50
Ω
SMA termination to terminate the CH 1 Markers M1 –
connector on the instrument.
Figure 2-17: Equipment connection for verifying the marker high and low level
accuracy
AWG70000A Series and AWGSYNC01 Technical Reference
2–31