Brief procedures
8.
Click the
Reset to default setup
button in the toolbar.
9.
Set the function generator output to square wave, 1 kHz, 5 V
p-p
.
10.
Turn on the output of the function generator.
11.
On the instrument, load the
PV_Square.wfmx
waveform as an output
waveform. Follow the steps below:
a.
In the Waveform List window, click
Open Waveform
and navigate to
C:\Program Files\Tektronix\AWG70000\Samples\PV
.
b.
Select
Open File
.
c.
In the Waveform List window, select (drag and drop) the
PV_Square.wfmx
waveform on the work space area.
12.
In the Run tab on the bottom of the screen, select
Triggered
, and set the
Trigger Input to
A
.
13.
Click the
CH1 Enable
button.
14.
In the Setup tab, click
Trigger
, and set the trigger level to 1.0 V, Rising, 50
Ω
.
15.
Click the
Play
button on-screen or on the instrument.
16.
Press the
All Outputs
button on the instrument to output the waveform.
17.
Verify that the output is displayed on the AWG work space.
18.
Verify that the output is displayed on the oscilloscope.
19.
Repeat the test for the Trigger B input: Move the cable from the Trigger A
input to the Trigger B input, and then under the Run tab, set the trigger input
to
B
.
20.
Verify that the Trigger B output is displayed on the AWG work space.
21.
Verify that the output is displayed on the oscilloscope.
22.
Disconnect the test setup.
2–20
AWG70000A Series and AWGSYNC01 Technical Reference
Содержание AWG70001A
Страница 2: ......
Страница 12: ...Service safety summary viii AWG70000A Series and AWGSYNC01 Technical Reference...
Страница 14: ...Preface x AWG70000A Series and AWGSYNC01 Technical Reference...
Страница 15: ...Specifications...
Страница 16: ......
Страница 42: ...AWGSYNC01 1 26 AWG70000A Series and AWGSYNC01 Technical Reference...
Страница 43: ...Performance verification...
Страница 44: ......