![Tektronix 502A Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/tektronix/502a/502a_instruction-manual_1078909051.webp)
Calibration
Procedure—
Type
502A
SENSITIVITY
Initial
Checking
Min. Amp.
Setting
Freq
.
and
Freq.
for
Proper
Amplitude
Response
.2
VOLTS
PER CM
4
cm
at 1 kc
1
mc
2.8
cm
100
mVOLTS
PER CM
4
cm
at 1 kc
400
kc
2.8
cm
50 mVOLTS
PER CM
4
cm
at 1 kc
400
kc
2.8
cm
20 mVOLTS
PER CM
4
cm
at
1
kc
200
kc
2.8
cm
10 mVOLTS
PER CM
4
cm
at 1 kc
200
kc
2.8
cm
5
mVOLTS
PER
CM
4
cm
at 1 kc
200
kc
2.8
cm
2
mVOLTS
PER
CM
4
cm
at 1 kc
200
kc
2.8
cm
1
mVOLTS PER
CM
4
cm
at 1 kc
200
kc
2.8
cm
.5 mVOLTS PER
CM
4
cm
at 1
kc
100
kc
2.8
cm
.2
mVOLTS
PER CM
4
cm
at 1 kc
100
kc
2.8
cm
.1
mVOLTS PER CM
4
cm at
1
kc
50
kc
2.8
cm
Remove
input signal from oscilloscope.
27.
Adjusting A
Input Attenuator Compensa
tions
Apply
1-kc square
waves from the Type 105 to input con
nector A
of
the vertical amplifier through the Input Time
Constant Standardizer.
The cable used should be terminated
into
the proper resistance on each
end.
Place
the
Input Selector switch in the A-DC position and
the SENSITIVITY
switch
in
the .2 VOLTS PER CM position.
Adjust
the
output of the Type
105 for
approximately 3 cm
of vertical
deflection.
Set the TRIGGER SELECTOR to
trigger off
the proper
beam, the TIME/CM to
.5 mSEC and the TRIGGERING
LEVEL
to trigger
the signal.
Adjust C410 for
an undistorted
square
wave.
Place
the SENSITIVITY switch
in the 2 VOLTS PER CM
position and
adjust
the Type 105 output for approximately
3
cm of
vertical deflection. Adjust C406A and C406C for
an
undistorted
square
wave.
Place
the
SENSITIVITY switch
in the 5 VOLTS
PER CM
position
and adjust
the Type 105 output for approximately
3
cm
of vertical deflection. Adjust C406G and C406J for
an
undistorted square
wave.
28.
Adjusting B
Input Attenuator Compensa
tions
Connect 1-kc square
waves from the Type 105 to input
connector
B of the vertical amplifier through the input Time
Constant
Standardizer. Place the Input Selector switch in the
B-DC
position and
the SENSITIVITY switch in the .2 VOLTS
PER
CM
position. Set the Type 105 output amplitude for
approximately 3
centimeters of vertical deflection. Adjust
C440 for
an undistorted square wave.
Place
the SENSITIVITY switch
in the 2 VOLTS PER CM
position and
adjust
the Type 105 output for approximately
3
cm of
vertical deflection. Adjust C407A and C407C for
an
undistorted square wave.
Place
the SENSITIVITY
switch in the 5 VOLTS PER CM
position
and adjust the Type 105 output for approximately
3
cm of
vertical
deflection. Adjust C407G and C407J for
an
undistorted
square
wave.
5-9
Содержание 502A
Страница 4: ......
Страница 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Страница 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Страница 30: ...Circuit Description Type 502A 3 12 NORMALLY CLOSED LOCATED ON REAR PANEL Fig 3 6 Simplified CRT Circuit ...
Страница 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Страница 34: ...NOTES ...
Страница 56: ...NOTES ...
Страница 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Страница 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Страница 67: ...Parts List Type 502A TOP 6 11 ...
Страница 70: ...Parts List Type 502A BOTTOM 6 14 ...
Страница 73: ...Parts List Type 502A REAR 6 17 ...
Страница 89: ...TYPE 502A OSCILLOSCOPE A ...
Страница 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Страница 91: ... TYPE 502A OSCILLOSCOPE A ...
Страница 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Страница 93: ......
Страница 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Страница 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Страница 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Страница 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Страница 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Страница 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...