![Tektronix 502A Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/tektronix/502a/502a_instruction-manual_1078909046.webp)
Calibration
Procedure
—
Type
502A
Fig.
5-4. Adjustment
of the GEOM
control. In waveform A the GEOM control
is
improperly set causing bowing of the vertical lines. In
waveform
B the
bowing of the vertical
lines has been reduced to a minimum by the proper adjustment of the GEOM control.
vertical
portions
of
the markers are displayed on the face
of
the
crt and adjust the TRIGGERING
LEVEL Control to
obtain
a
stable
display. Set the
GEOM control for minimum
curvature
of the vertical
lines at the sides of the crt.
8. Adjusting CRT
HORIZ SENS
BAL and
HORIZ
BEAM
REGIS
From
the Type 180A apply 100 MICROSECONDS
and 1
MILLISECONDS
markers to the A input of the LOWER BEAM
and
the B input of the UPPER BEAM. Switch the UPPER
BEAM
Input
Selector to B-DC and the SENSITIVITY switches
of
both beams to 5 VOLTS PER CM. With the Vertical
POSITION Controls,
position the traces until the markers of
the
LOWER
BEAM are just touching the markers of
the
UPPER BEAM.
Adjust the
HORIZ BEAM REGIS until the
first and ninth
time markers are
as near being above one another as
possible.
Now adjust
the CRT HORIZ SENS BAL until the
sensitivity of
the two beams is the same. At this point in
the
calibration the two beams are being matched to each
other and
the
time markers may not coincide with the
graticule
marks.
NOTE
Every
time the CRT HORIZ SENS BAL is adjusted
the INT
BAL
must be checked. The HORIZ BEAM
REGIS
may
also have to be readjusted at this time.
The
above
three
controls interact with each other.
9.
X1 CAL
and
X20
CAL Adjustments
Set
the
UPPER BEAM
Input Selector
to the A-DC position.
With
1
MILLISECONDS and
100 MICROSECONDS time
markers
applied
to
the A input of each beam reset the
HORIZONTAL
DISPLAY switch to X20 position. Adjust
the
X20
CAL for a display of one 100 MICROSECONDS time
marker
per 2 cm.
Now turn the HORIZONTAL DISPLAY
Switch
to X1 and adjust the X1 CAL for one 1 MILLI
SECONDS
time marker every cm.
Fig.
5-5.
Adjustment of the X20 control. The X20 CAL control is
adjusted
in
the
calibration procedure
for the correct magnified
sweep
rate. The control should be set
so that markers occur at
2
centimeter intervals.
There
will
be
interaction between the X1 and X20 adjust
ments.
After this interaction has been removed
the X2 and
X5 and
X10 positions
of the HORIZONTAL DISPLAY switch
should
be checked. Disconnect the
time markers.
5-4
Содержание 502A
Страница 4: ......
Страница 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Страница 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Страница 30: ...Circuit Description Type 502A 3 12 NORMALLY CLOSED LOCATED ON REAR PANEL Fig 3 6 Simplified CRT Circuit ...
Страница 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Страница 34: ...NOTES ...
Страница 56: ...NOTES ...
Страница 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Страница 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Страница 67: ...Parts List Type 502A TOP 6 11 ...
Страница 70: ...Parts List Type 502A BOTTOM 6 14 ...
Страница 73: ...Parts List Type 502A REAR 6 17 ...
Страница 89: ...TYPE 502A OSCILLOSCOPE A ...
Страница 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Страница 91: ... TYPE 502A OSCILLOSCOPE A ...
Страница 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Страница 93: ......
Страница 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Страница 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Страница 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Страница 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Страница 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Страница 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Страница 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Страница 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Страница 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...
Страница 110: ...TYPE 502A TENT S N 23430 PARTS LIST CORRECTION CHANGE TO V152 157 0104 01 Selected 6AL5 M913U 665 ...