![Tektronix 502A Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/tektronix/502a/502a_instruction-manual_1078909047.webp)
Calibration
Procedure
—Type
502A
10.
NORM MAG REGIS
Turn
the TRIGGERING
LEVEL to RECURRENT and place
the
HORIZONTAL DISPLAY switch
in the X20 position. Posi
tion
the start of
the
upper trace under the
vertical centerline
of
the graticule.
Rotate the HORIZONTAL DISPLAY switch
to
NORMAL
and adjust the NORM MAG REGIS Control
so
that
the start of
the sweep again lies under the vertical
centerline
of
the graticule.
Repeat adjustments 9 and 10
until
both are set correctly.
11.
Check Timing from .1 millisecond
through
5
seconds
From
the
Type 180A apply time markers to the A input of
both
beams
as
indicated below. Adjust for
a stable display
and
check for proper display.
Disconnect
Time Markers.
TIME/CM
TIME
MARKERS
DISPLAY
1
mSEC
1
MILLISECONDS
1
marker/cm
2
mSEC
1
MILLISECONDS
2
markers/cm
5
mSEC
5
MILLISECONDS
1
marker/cm
10
mSEC
10
MILLISECONDS
1
marker/cm
20 mSEC
10
MILLISECONDS
2
markers/cm
50
mSEC
50
MILLISECONDS
1
marker/cm
100
mSEC
100
MILLISECONDS
1
marker/cm
200
mSEC
100
MILLISECONDS
2
markers/cm
500
mSEC
500
MILLISECONDS
1
marker/cm
1
SEC
1
SECONDS
1
marker/cm
2
SEC
1
SECONDS
2
markers/cm
5
SEC
5
SECONDS
1
marker/cm
.1
mSEC
100
MICROSECONDS
1
marker/cm
12.
Fast Sweep Rate Adjustments
Turn
the
TIME/CM Control
to 10 μSEC and the TRIGGER
ING
LEVEL to RECURRENT. Set the input switch of the test
oscilloscope
to DC.
Connect a
probe from the test oscilloscope to pin 3 of
V183 and adjust
C311 for a flat base on the holdoff
waveform. This
is just
a preliminary
adjustment of C311.
Adjustment
of the faster
sweep rates is indicated in the
chart
below. The
1 μSEC and the 5 μSEC timing adjustments
will
interact and therefore will have
to be rechecked
several
times.
Apply
time markers
to both A
inputs and trigger the
display.
Disconnect
Time Markers.
TIME/CM
TIME
MARKERS
FROM TYPE
180A
ADJUST
OR
CHECK
10 μ
SEC
10
MICROSECONDS
Adjust
C160C for
1
marker/cm
20 μSEC
10
MICROSECONDS
Check for
2
markers/cm
50 μSEC
50
MICROSECONDS
Check
for 1
marker/cm
1
μ
SEC
1
MICROSECONDS
Adjust
C311
for
1
marker/cm
2
μ
SEC
1
MICROSECONDS
Check
for 2
markers/cm
5
μ
SEC
5
MICROSECONDS
Adjust
C160A for
1
marker/cm
13. EXT HORIZ AMP DC
BAL
Rotate
the
HORIZONTAL DISPLAY switch
through the
EXT
VOLTS/CM
positions and adjust the
EXT HORIZ AMP DC BAL
Control so
that
there is no horizontal shift of the spots as
the HORIZONTAL DISPLAY switch
is rotated.
14.
Checking External
Gain and Attenuator
Apply
5 volts
to
the EXTERNAL connector from the AMPLI
TUDE
CALIBRATOR.
With
the HORIZONTAL DISPLAY switch
set
in the 2 EXT. VOLTS/CM position check for a deflection
of 2.5
cm.
Check
the other positions of the display
switch
for
the proper deflection. Return the HORIZONTAL DISPLAY
switch
to
the NORMAL position
and disconnect the jumper.
NOTE
The
following adjustments apply to both of the
vertical
amplifiers. Complete the calibration
of one
amplifier
before proceeding to the other.
A.
B.
C.
Fig.
5-6. Adjustment
of C311. Waveforms A and
C result from improper settings of C311. Capacitor C311 should be adjusted
so that
waveform
B
is obtained on the test oscilloscope.
5-5
Содержание 502A
Страница 4: ......
Страница 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Страница 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Страница 30: ...Circuit Description Type 502A 3 12 NORMALLY CLOSED LOCATED ON REAR PANEL Fig 3 6 Simplified CRT Circuit ...
Страница 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Страница 34: ...NOTES ...
Страница 56: ...NOTES ...
Страница 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Страница 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Страница 67: ...Parts List Type 502A TOP 6 11 ...
Страница 70: ...Parts List Type 502A BOTTOM 6 14 ...
Страница 73: ...Parts List Type 502A REAR 6 17 ...
Страница 89: ...TYPE 502A OSCILLOSCOPE A ...
Страница 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Страница 91: ... TYPE 502A OSCILLOSCOPE A ...
Страница 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Страница 93: ......
Страница 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Страница 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Страница 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Страница 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Страница 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Страница 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Страница 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Страница 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Страница 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...
Страница 110: ...TYPE 502A TENT S N 23430 PARTS LIST CORRECTION CHANGE TO V152 157 0104 01 Selected 6AL5 M913U 665 ...