Operators Familiarization
Externa! Triggering
Any of the four vertical channels in the 2252 can be used as a source of A and
B trigger signals. When you need a trigger signal source different from the
one derived from displayed signals, you can use any free vertical input
channel. CH 1 and CH
2
can condition a wide range of signals to produce
triggers over the full vertical deflection range from millivolts to hundreds of
volts. CH 3 and CH 4 have two basic attenuation factors {0.1 and 0.5 volts per
division), making them especially useful for triggering on and viewing digital
signal levels.
Auto Setup
Pressing the AUTO SETUP button automatically sets up the front-panel
controls based on the characteristics of the applied signal. The voltage ampli-
tude, sweep settings, trigger parameters, vertical and horizontal positioning,
and trace intensities are preset to produce a usable waveform display. The
waveform is horizontally centered and vertically positioned within the crt dis
play. The voltage and frequency characteristics of the input signal must be
within the limits of the 2252 specifications given in Section 6.
Auto-setup action for each front-pane! control is shown in Appendix A.
Minimum Setup
Pressing and holding the AUTO SETUP button until the instrument returns to
normal operation initiates the minimum-setup function. This automatically
sets up the front-pane! controls but leaves many of the controls unchanged
from what the user has set.
Minimum setup action for each front-panel control is shown in Appendix A,
Remote Control via the Communications
Port
i
When connected in a system with a host computer or system controller, the
2252 can be operated by the controller. The front panel controls and measure
ments are accessible via the communications interface. Digitized waveforms
and measurement results can be gathered by the remote controller for tong
term data collection and /or analysis using system programming. The stan
dard communications interface is a GPIB (general purpose interface bus) for
parallel byte-wide data transfer. The GPIB commands understood by the in
strument and programming hints are found in Appendix B.
2252 Operators
3-5
Содержание 2252
Страница 17: ...SECTION 1 INTRODUCTION 2252 Operators ...
Страница 18: ......
Страница 26: ......
Страница 27: ...SECTION 2 CONTROLS CONNECTORS AND INDICATORS 2252 Operators ...
Страница 28: ... f ...
Страница 55: ...SECTION 3 OPERATORS FAMILIARIZATION 2252 Operators ...
Страница 56: ......
Страница 107: ...SECTION 4 OPERATOR CHECKS AND ADJUSTMENTS 2252 Operators ...
Страница 108: ......