Controls, Connectors, and Indicators
(4 7 )
SET
MEAS’MT
CHANN
EL—Displays
menus
for
setting
measurement-source channels for active measurement modes.
There are two one-page menus available; one for voltmeter and rise/falt
measurements, and one for single-channel Counter/Timer measure
ments. Two-channel measurements, such as delta time, frequency
ratio, and propagation delay, have two-page menus. Each menu lists
all valid input channels that can be selected for that measurement
mode.
If the SET MEAS’MT CHANNEL button is pressed for an invalid mode,
one of the following messages will be displayed for two seconds:
1. SELECT A MEASUREMENT-When no measurement mode is
active.
2. NO MEAS CHANNEL NEEDED—When a selected measure
ment mode (such as cursor time) does not require a measure
ment channel to be set.
When a SET MEAS’MT CHANNEL menu is displayed, changing the
Horizontal MODE, except between ALT and B, clears the menu.
(48) LAST MEAS’MT—
Recalls the last active measurement mode to the
display and resets the measurement channel. If the last active measure
ment mode is already displayed when the LAST MEAS’MT button is
pressed, only the measurement channel is reset.
The measurement channel for VOLTMETER, CURSOR VOLTS, and
RISE o r FALL measurement modes is set to CH 1, if CH 1 is displayed,
otherwise to CH 2. If neither CH 1 nor CH 2 is displayed, CH 1 will be
turned on and will be the measurement channel.
The measurement channel for single-channel Counter/Timer
measurements is set to the lowest number displayed channel (may be
set to CH 1, CH 2, CH 3, CH 4, or ADD).
.
Measurement channels for time measurement modes (ALT and B Hori
zontal modes) and dual-channel Counter/Timer measurements are set
to the lowest number displayed channel for the delay time (or B source
1) and the next lowest number displayed channel for the delta-delay
time, (or B source 2} if more than one channel is displayed. Both are set
to the same channel when only one channel is displayed. CH 1 is con
sidered the lowest numbered channel, and ADD is considered the
highest numbered channel.
2252
Operators
2-23
Содержание 2252
Страница 17: ...SECTION 1 INTRODUCTION 2252 Operators ...
Страница 18: ......
Страница 26: ......
Страница 27: ...SECTION 2 CONTROLS CONNECTORS AND INDICATORS 2252 Operators ...
Страница 28: ... f ...
Страница 55: ...SECTION 3 OPERATORS FAMILIARIZATION 2252 Operators ...
Страница 56: ......
Страница 107: ...SECTION 4 OPERATOR CHECKS AND ADJUSTMENTS 2252 Operators ...
Страница 108: ......