Operator Checks and Adjustments
2.
Connect the probe tips to the PROBE ADJUST connector and the probe
ground leads to scope ground.
3.
Set:
1
Vertical MODE
CH 1 & CH 2
Horizontal MODE
A
4.
Press AUTO S £ fU P button.
5.
Set the CH 1 VOLTS/DIV setting to 0.1 V {10 mV with probe discon
nected) and vertically center the PROBE ADJUST square-wave signal.
6.
Check the square-wave signal for overshoot and rolloff {see Figure 4-1).
If necessary, use the special adjustment too! supplied in the probe
accessory package to adjust the low-frequency compensation for a
square front corner on the square wave displayed.
7.
Press the CH 2 Vertical MODE button to turn CH
2
on in the display, and
press the CH 1 Vertical MODE button to remove the CH 1 trace from the
display.
Figure 4-1 . Probe compensation.
4-4
2252
Operators