Operator Checks and Adjustments
Timing Checks
The time measurement cursors may be used to check the horizontal deflec
tion system.
1. Preset instrument controls and obtain a baseline trace and set the
A SEC/DIV switch to 0.1 ms. Vertically center the baseline trace.
-T
2. Press the TIME MEASUREMENTS button to call up the Time Measure
ment Mode menu on the crt and select
SEC~*I function for measuring
time difference by pressing the Menu Select button opposite menu label.
3. Align the reference cursor to the second vertical graticule line using the
H- OR DELAY TIME control (both cursors are positioned together).
4. Adjust the
control for a reading of 800.0 fxs.
5. Check that the cursors are eight divisions apart.
6. Press the CLEAR DISPLAY button to remove the cursors from the
display.
4 -6