CH 2: Obtain a free-running trace with no signal applied to the
vertical channel input and CH 1 POSition control set to the OFF
detent. While switching the CH 2 VOL TS/DIV switch between 50 m
and 1 m, adjust the CH 2 STEP ATTEN BAL for minimum trace shift
between adjacent positions.
Horizontal Gain
Set the SEC/DIV switch to X-Y. Connect the CH 1 probe tip to
an accurate 0.2 volt source (such as a TEKTRONIX 067-0502-01
Standard Amplitude Calibrator). Set the CH 1 VOL TS/DIV switch to
50 m and the CH 2 POSition control to midrange. Adjust the HORIZ
GAIN adjustment for exactly four horizontal divisions of deflection
between dots.
Horizontal Timing
Connect the CH 1 probe tip to some accurate timing standard
(such as the TEKTRONIX 2901 Time-Mark Generator). With the
SEC/DIV switch set to 1 m, apply one millisecond time marks.
Adjust the Trigger controls for a stable display. Adjust the SWP CAL
adjustment for exactly eight divisions of deflection between the
second and tenth time mark (one time mark/division).
22
212 Operators Handbook