1730–Series Checks and Adjustments
5–21
a. Connect a 50
Ω
precision cable and 50 to 75
Ω
minimum loss pad from
the leveled sine wave to the CH–A input. Teminate the loop–through in
75
Ω
.
b. Select EXT REF.
c.
Set the leveled sine wave generator to 50 kHz (reference frequency) and
set the output amplitude for a 100 IRE (0.7 V) display amplitude.
d. CHECK – the flat response using the 50 kHz response as a reference.
Changing the frequency of the sine wave generator, check that the
response is within
±
2% from 250 kHz to 6 MHz.
e.
Change the 1730–Series VERTICAL GAIN to X5.
f.
Set the leveled sine wave generator to 50 kHz and set the output
amplitude for a 100 IRE (0.7 V) display amplitude.
g. CHECK – flat response using the 50 kHz response as a reference.
Check that the amplitude is 100 IRE (0.7 V)
±
5% from 250 kHz to
6 MHz.
h. Move leveled sine wave generator output to CH–B and terminate the
remaining side of the loop–through INPUT with a 75
W
termination.
i.
Select CH–B INPUT.
j.
CHECK – flat response using the 50 kHz response as a reference.
Check that the amplitude is 100 IRE (0.7 V)
±
5% from 250 kHz to
6 MHz.
k. Turn off X5 VERTICAL GAIN.
l.
CHECK – the flat response using the 50 kHz response as a reference.
Check that the response is within
±
2% from 250 kHz to 6 MHz.
14. Check PIX MON OUT Frequency Response
REQUIREMENT – PIX MON OUT Frequency Response with a 50 kHz
reference; within 3% from 250 kHz to 6MHz.
a. Connect a coaxial cable to the + input on the Peak To Peak Detector
Amplifier and connect the other end of the cable to the Peak to Peak
Detector Head.
b. Press the + Input button on the Peak To Peak Detector Amplifier.
c.
Adjust the Peak to Peak Detector Amp Input Level control fully
clockwise.
Содержание 1730 Series
Страница 4: ......
Страница 12: ...Contents viii...
Страница 17: ...Introduction and Specifications...
Страница 18: ......
Страница 32: ...1730 Series Introduction 1 14...
Страница 33: ...Operating Instructions...
Страница 34: ......
Страница 62: ...1730 Series Operating Instructions 2 28...
Страница 64: ...Service Safety Summary S 2 1730 Series B070000 Above...
Страница 66: ......
Страница 67: ...Installation...
Страница 68: ......
Страница 82: ...1730 Series Installation 3 14...
Страница 83: ...Theory of Operation...
Страница 84: ......
Страница 115: ...Checks and Adjustments...
Страница 116: ......
Страница 159: ...Maintenance...
Страница 160: ......
Страница 180: ...1730 Series Maintenance 6 20 3 Remove the board by slipping it through the front panel opening...
Страница 184: ...1730 Series Maintenance 6 24 Figure 6 8 Repackaging a 1730 Series instrument...
Страница 185: ...Options...
Страница 186: ......
Страница 189: ...Replaceable Electrical Parts...
Страница 190: ......
Страница 222: ...1730 Series Replaceable Electrical Parts 8 32...
Страница 223: ...Diagrams Circuit Board Illustrations...
Страница 224: ......
Страница 246: ......
Страница 247: ...Replaceable Mechanical Parts...
Страница 248: ......
Страница 255: ...1730 Series Waveform Monitor FIG 1 EXPLODED VIEW A1 A3 A2 A3A1 A10 A11...
Страница 256: ...1730 Series B070000 Up...
Страница 257: ......
Страница 258: ......