GENERAL 

TP3005DM is a high performance and precision DC regulated power supply. TP3005DM has constant 

voltage mode, constant current mode, over-voltage protection function, overload protection, and 8 hours 

continuous full load features. Voltage and current value can be adjusted linearly. With the extract 

stability and enhanced responsibility, this power supply is suitable for bench, laboratory, university, high 

school, and enterprise use or where needed a high performance and precision regulated DC power 

supply. 

 

 

3. FEATURES AND SPECIFICATIONS

 

 

3.1 OPERATION CONDITION 

 

Input Voltage

 

110V AC or 220V AC

±

10% 60/50Hz 

Power Consumption

 

300W 

Operating Condition

 Temperature

0

40

 

Relative Humidity*

 90%RH 

Storage Condition

 Temperature

-20

℃~

80

 

Relative Humidity*

 80%RH 

*Non-condensing 

 

3.2 TECHNICAL SPECIFICATION 

 

Constant Voltage Mode (CV) 

Voltage Range: 

0~30V (main output), fixed 5V (USB output) 

Line Effect: 

1×10

-4 

+3mV (main output) 

100mV (USB output) 

Loading Effect: 

1×10

¯4

+3mV (output current

3A) (main output) 

2×10

¯4

+3mV (3A

output current

5A) (main output) 

100mV (USB output) 

Noise and Ripple: 

1mV (main output) 

2mV (USB output) 

Temperature Coefficient 

150ppm/

 

Constant Current Mode (CC) 

Current Range 

0~5A (main output), Max. 2A (USB output) 

Line Effect: 

2×10

-3

 +3mA (main output) 

Loading Effect: 

1×10

¯3

+3mA (output current

3A) (main output) 

2×10

¯3

+3mA (3A

output current

5A) (main output) 

Temperature Coefficient: 

500ppm/

 

Содержание TP-3005D

Страница 1: ...IONS 4 3 1 OPERATION CONDITION 4 3 2 TECHNICAL SPECIFICATION 4 4 OPERATING INSTRUCTIONS 5 4 1 INSTRUMENT DESCRIPTION 5 4 1 1 FRONT PANEL DESCRIPTION 5 4 1 2 REAR PANEL DESCRIPTION 6 5 POWER SUPPLY DES...

Страница 2: ...z For your own safety and that of the instrument you must follow the procedure described in this instruction manual z Do not use this instrument in a location where there is explosive gas in the vicin...

Страница 3: ...ghly adjustment or VOLTAGE FINE CURRENT FINE tune knob for fine adjustment Once the voltage or current value is reached to desire value appear on Display Panel 1 2 DURING USE 1 Ensure the voltage and...

Страница 4: ...0V AC or 220V AC 10 60 50Hz Power Consumption 300W Operating Condition Temperature 0 40 Relative Humidity 90 RH Storage Condition Temperature 20 80 Relative Humidity 80 RH Non condensing 3 2 TECHNICAL...

Страница 5: ...R When the power supply is at the constant current mode this LED light will be on 5 CV constant voltage mode INDICATOR When the power supply is at the constant voltage mode this LED light will be on 6...

Страница 6: ...t value 12 main positive terminal 0 30V positive output 13 GND grounding terminal This terminal is connecting to the casing and the Earth 14 main negative terminal 0 30V negative output 15 USB output...

Страница 7: ...E COARSE TUNE KNOB 8 and VOLTAGE FINE TUNE KNOB 9 to adjust and fine turn the voltage to give a desired output voltage 5 Connect the circuit to the MAIN TERMINAL 12 14 6 When the CC INDICATOR 4 is on...

Страница 8: ...INDICATOR 5 will off and the CC INDICATOR 4 will on 6 Adjust the CURRENT KNOB 10 11 to give a desired output current 7 Repress the ACCIDENTAL PROTECTION KEY 2 to cut off the output 8 Then the CC INDIC...

Страница 9: ...clockwise to open the fuse socket 5 Replace the fuse with identical rating Rating see section 4 1 1 6 Use the cross screw drive and turn clockwise to close the fuse socket Caution z Ensure no power i...

Страница 10: ...ical rating fuse When operate in CV mode the voltage drop down and the CC indicator on It is the current protect The desired current range is under the circuit gain The power supply switched to CC mod...

Отзывы: