10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • [email protected]
2
GENERAL INFORMATION
REPEATABILITY (1 week at constant ambient temperature):
K, J, T, E, N;
±0.2°F typ.
B, R, S, G, C, D;
±1.0°F typ.
RTD, THERMISTOR;
±0.1 °F typ.
OHMS;
±0.01 % rdg typ.
METER-MODE INPUT-CURRENT (THERMOCOUPLES):
50nA typ.
METER-MODE READING-RATE:
1 reading/second typ.
METER-MODE EXCITATION-CURRENTS:
100
Ω
RTD;
<2mA
1000
Ω
RTD;
<200
μ
A
999.99
Ω
;
<2mA
9999.9
Ω
;
<200
μ
A
99999
Ω
;
<50
μ
A
THERMISTOR; T
≤
113°F <50
μ
A
T> 113°F, <2mA
CALIBRATOR MODE OUTPUT LOAD (THERMOCOUPLES):
500
Ω
min.
CALIBRATOR-MODE EXCITATION-CURRENTS:
100
Ω
RTD;
0.5 - 1mA (0.1 - 2mA typ.)
1000
Ω
RTD;
50 - 100
μ
A (10-200
μ
A typ.)
999.99
Ω
;
Same as 100
Ω
RTD
9999.9
Ω
;
Same as 1000
Ω
RTD
99999
Ω
;
10 - 20
μ
A (1 - 20
μ
A typ.)
THERMISTOR;
T
≤
113°F, 10 - 20
μ
A
T> 113°F, 0.5- 1mA
MAXIMUM COMMON MODE VOLTAGE
: 42V peak to earth.
KEYPAD
: 12 momentary switches with tactile feedback select;
• On/Off
• Change/Enter
• Clear/-0
• Sensor Select/1
• °F/°C/2
• Resolution/3
• Operate/4
• Store/5
• Recall/6
• Calibrator/Meter/Up Ramp/7
• Calibrator Step/Down Ramp/8
• Calibrator Ramp/Meter Hold/9