GENERAL INFORMATION
Figure 6. 4-Wire Resistance Measurement
5. DATA LOGGING AND RECALL
Operating as a meter, the unit can save and recall up to 10 readings (each with independent
parameters: eg. °F/°C, TC-type and resolution) in random access memory.
To Save Readings:
1. While in the METER OPR mode, depress the STORE key. Note that the OPR annunciator is replaced
by STO.
2. Next depress one of the memory location numbers (keys 0, 1, ..., 9). The corresponding
memory annunciator turns on and the meter returns momentarily to the OPR mode, displaying
thermocouple temperature.
eg. STORE 7
NOTES:
• The enabled location annunciator indicates that the meter reading
at the instant that the location number key was pressed, was stored in
memory.
• When data is stored in a location already filed with data, the new data
replaces the old.
• Stored readings are retained during power-off.
To Recall Readings:
1. While in the METER OPR mode, readings stored earlier can be displayed by first depressing the
RECALL key. Note that the OPR annunciator is replaced by RCL, and all memory location