SERVICE INFORMATION
10. Press the ENTER function button. The "0.0 °F" display will stop blinking, and the "Ω" will
continue to blink. You are now in the copper mode. If or when you turn off the power (or
remove the battery) the unit will no longer be in the copper mode and you would have to
repeat the above steps to return to the copper mode. As long as you see the "Ω" blinking
you are in the copper mode.
11. While in the copper mode attach the Copper Mode Calibrator Verification Cable to the
connector on the top of the 845 that is labeled T1. Attach the other end of the cable to a
precision voltmeter.
12. Press the CHANGE function button. The "0.0 °F" display will blink and "NUM" will appear on
the display.
13. Press 3 2 0 buttons and then press the ENTER function button. "NUM" will disappear and
"32.0 °F" will appear, not blinking. The voltmeter should read 0.000 mV.
14. Press the CHANGE function button. The "32.0 °F' display will blink and "NUM" will appear on
the display.
15. Press 2 5 0 0 0 buttons and then press ENTER function button. "NUM" will disappear and
"2500.0 °F" will appear, not blinking. The voltmeter should read 54.856 mV.
16. Press the CHANGE function button. The "2500.0 °F" display will blink and "NUM" will appear
on the display.
17. Press 3 2 0 buttons and then press the ENTER function button. "NUM" will disappear and
"32.0 °F" will appear, not blinking. The voltmeter should read 0.000 mV.
18. Press the SENSOR SELECT function button. The sensor type will change to type J. The "K"
will disappear from the display and a "J" will appear on the display. The voltmeter should read
0.000 mV.
19. Press the SENSOR SELECT function button. The sensor type will change to type T. The "J"
will disappear from the display and a "T" will appear on the display. The voltmeter should read
0.000 mV.
20. Press the SENSOR SELECT function button. The sensor type will change to type E. The "T"
will disappear from the display and an "E" will appear on the display. The voltmeter should
read 0.000 mV.
21. Press the SENSOR SELECT function button. The sensor type will change to type N. The "E"
will disappear from the display and a "N" will appear on the display. The voltmeter should read
0.000 mV.
22. Press the SENSOR SELECT function button. The sensor type will change to type B. The "N"
will disappear from the display and a "B" will appear on the display. "LO" will also appear on
the display.
23. Press the CHANGE function button. The "32.0 °F" display will blink and "NUM" will appear on
the display.
24. Press 1 0 0 00 buttons and then press ENTER function button. "NUM" will disappear and
"1000.0 °F" will appear, not blinking. The voltmeter should read 1.438 mV.