10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • [email protected]
29
SERVICE INFORMATION
Table 4: Calibration of Calibrator-Mode Resistance Functions (RTD-100/1000)
STEP SENSOR
SELECT
UNIT
OUTPUT
CURRENT
SOURCE INPUT
ADJUST
VERIFY
DMM READING*
1
RTD-100
1562.0
°
F
1mA
P38
0.39048V
2
RTD-100
32.0
°
F
1mA
√
0.10000±0.00003V
3
RTD-100
-328.0
°
F
1mA
√
0.01852±0.00006V
4
RTD-1000 1562.0
°
F
100µA
P51
0.39048V
5
RTD-1000
32.0
°
F
100µA
√
0.10000±0.00003V
6
1,000Ω
900.00Ω
1mA
√
0.90000±0.00010V
7
1,000Ω
1.00Ω
1mA
√
0.00100±0.00010V
8
10,000Ω
9000.0Ω
100µA
√
0.900000±0.00010V
9
10,000Ω
10.0Ω
100µA
√
0.000100±0.00010V
* exclusive of noise.
Table 5: Calibration of Meter-Mode Resistance Functions (RTD-Thermistor)
STEP
SENSOR
SELECT
UNIT
OUTPUT
ADJUST
VERIFY
DMM READING*
1
RTD-100
390.48Ω
P1
1562.0
°
F ±0.6
2
RTD-100
100.00Ω
√
32.0±0.2
°
F
3
RTD-100
18.52Ω
√
-328.0±0.5
°
F
4
1,000Ω
900.0Ω
P1
900.00Ω
5
1,000Ω
1.00Ω
√
1.00±0.20Ω
6
100,000Ω
90,000Ω
P1
90,000Ω
7
100,000Ω
100Ω
√
100±20Ω
8
Thermistor
47.0Ω
P1
293.0
°
F
9
Thermistor
2252Ω
√
77.0±0.2
°
F
10
Thermistor
75,790Ω
√
-40.0±0.2
°
F
* exclusive of noise.