3. Electrical Connections
14
3.2
E
LECTRICAL INSTALLATION OF CURRENT OUTPUT
Below is reported the procedure for the electrical installation of the probes
equipped with the current output with two-wire connection, as shown in Figure
8
1.
Rotate the upper element (3) of the sensor counterclockwise and
extract the upper element.
2.
Loosen the locking nut (2) of the threaded connector and pass the
measuring cable (1) through it.
3.
Uncover the ends of the cable by 0.5 cm, apply the cable lugs (
ϕ
max
=0.5 mm2) and connect them to the 2-conductor connection
respecting the polarity indicated on the terminal (4).
4.
Fully insert the upper element of the sensor (3) on the probe body (5)
and tighten it by turning clockwise until it stops.
5.
Insert the measurement cable as far as possible into the upper element
of the strain relief sensor.
6.
Tighten the lock nut (2) on the threaded connector.
Terminal
Function
Value
+
Power supply (+V
in
)
+12 ÷ 30 Vdc
-
Power supply (GND)
0