2. Installation
8
DO NOT TOUCH THE CAP AND ELECTRODES
Do not touch the membrane on the cap, nor the electrodes placed in the
lower part of the sensor; do not damage them and avoid that they come
into contact with greasy substances. Otherwise, the sensor will no longer
function accurately. In this case, replace the cap with a new one or send
the sensor to the manufacturer for cleaning the electrodes.
WARNING!
During all assembly, maintenance or other operations it is recommended not
to touch the membrane, in order to avoid damage that could lead to
malfunction of the probe.
2.3
INFORMATION ON THE ELECTROLYTE GEL
KEEP THE PRODUCT OUT OF REACH OF CHILDREN
!
E
LECTROLYTE GEL
.
DO NOT SHAKE THE BOTTLE BEFORE USE!
✓
The electrolyte is sensitive to oxidation: always keep the bottle of
electrolyte closed after use.
✓
Do not transfer the electrolyte into other containers.
✓
The electrolyte must not be stored beyond the permitted terms; for
the expiry date see the label on the electrolyte bottle.
✓
Pour in the electrolyte avoiding, where possible, the formation of
bubbles. The small air bubbles are not a problem, while the larger
bubbles rise towards the upper edge of the cap and affect the
measurement.
✓
Store the gel electrolyte bottle with the cap facing down, as indicated
on the bottle label.
Содержание NCL C10
Страница 2: ...VIA FRANCESCO ANTOLISEI 25 00173 ROME ITALY 39 06 4576 3210 www tsens biz info tsens eu ...
Страница 4: ......
Страница 8: ...1 Introduction 4 Figure 2 version with voltage output standard supply ...
Страница 31: ...7 Notes 27 7 NOTES ...
Страница 32: ...VIA FRANCESCO ANTOLISEI 25 00173 ROME ITALY 39 06 4576 3210 www tsens biz info tsens eu ...