2. Installation
9
WARNING
!
RISK DUE TO A DANGEROUS SUBSTANCE
.
When using hazardous substances, please note that updated safety data
sheets of the manufacturers of these substances are available. Since the risk
potential of a substance can be re-evaluated at any time based on new
knowledge, the safety data sheet must be checked regularly and replaced if
necessary.
The system operator is responsible for the availability of the updated version
of the safety data sheet and for the preparation of the risk
assessment of the workplaces concerned connected to it.
The information relating to the gel electrolyte and the safety
data sheet are available on the manufacturer's website at
http://www.aedes.info/prodotto.asp
2.4
M
OUNTING ON THE PROBE HOLDER
Insert the probe into a dedicated probe holder in the appropriate housing. A
distance of about 2 cm must be ensured between the membrane and the
bottom of the probe holder as indicated in Figure 6 on the next page, in order
to ensure optimal operation of the probe.
The phase of installing the probe in the probe holder must be carried out
delicately; do not push too hard to avoid the risk of damaging the membrane
due to excessive pressure. In order to avoid the occurrence of an overpressure
it is recommended to open at least one drain valve or sampling cock. Once the
probe is blocked in the probe holder at the desired height, it is possible to close
the drain valve or sampling cock, and gradually open the delivery cock.
Содержание NCL C10
Страница 2: ...VIA FRANCESCO ANTOLISEI 25 00173 ROME ITALY 39 06 4576 3210 www tsens biz info tsens eu ...
Страница 4: ......
Страница 8: ...1 Introduction 4 Figure 2 version with voltage output standard supply ...
Страница 31: ...7 Notes 27 7 NOTES ...
Страница 32: ...VIA FRANCESCO ANTOLISEI 25 00173 ROME ITALY 39 06 4576 3210 www tsens biz info tsens eu ...