![Tecnomed Italia MORPHEUS Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/tecnomed-italia/morpheus/morpheus_installation-use-and-maintenance-manual_1072005009.webp)
9
ENGLISH
or indirect property damage resulting from non-compliance with this requirement.persone
e/o danni a cose diretti o indiretti derivati dell’inosservanza di tale prescrizione.
ATTENTION! In order to ensure the functionality and safety of the product and to avoid any
wear-related damages, the product requires routine and extraordinary maintenance interventions
(specified in this manual) that must be carried out at specified intervals by your Tecnomed Italia
s.r.l. dealer. For more information, see the paragraph “Maintenance” on page 37.
DANGER! Do not disassemble the device. In case of malfunction, please contact an
authorised Tecnomed Italia s.r.l. centre.
TECNOMED ITALIA S.R.L. IS RESPONSIBLE FOR THE SAFE USE OF THE PRODUCT AND
GUARANTEES THE PRODUCT ONLY IF THE CONDITIONS PROVIDED IN THIS MANUAL
ARE RESPECTED.
For any request, always indicate the date of purchase, the model of the
device and the serial number. For information on technical support: [email protected]
3.1 - WORK ENVIRONMENT
FLOOR COMPOSITION
The quality of the floor must meet the capacity standards standards specified in DIN1055
sheet 3 and present the pressure resistance laid down in DIN 18560 T1.
ENVIRONMENTAL REQUIREMENTS:
•
Environmental temperature: from +10°C to + 40°C.
•
Optimum environmental temperature: from 15°C to 30°C.
•
Relative air humidity: from 30 to 75%.
•
Ambient pressure from 600 to 1060 hpa
3.2 - MAXIMUM LOADS
• The maximum allowed load on the surgical chair is 200 Kg. Do not exceed this value.
• The maximum allowed load on the sampling armrest is 10 Kg. Do not exceed this value.
• The maximum allowed load on the blood sampling armrest is 10 Kg. Do not exceed this value.
• The maximum allowed load on the thigh rests is 30 Kg. Do not exceed this value.
• The maximum allowed load on the headrest is 8 Kg. Do not exceed this value.
DANGER! Tecnomed Italia s.r.l. denies any liability, expressed or implied, and cannot
be held liable for injuries to persons and/or direct or indirect damages to property, deriving
from failure to comply with the indications above.
Содержание MORPHEUS
Страница 1: ...V 8 00 CODE DE1000X Installation use and maintenance manual MORPHEUS surgical chair ...
Страница 14: ...14 ENGLISH 5 3 1 DIMENSIONS ...
Страница 19: ...19 ENGLISH Positioning examples for standard configuration ...
Страница 20: ...20 ENGLISH ...
Страница 24: ...24 ENGLISH 7 8 9 10 11 12 ...
Страница 25: ...25 ENGLISH 13 14 15 X4 16 17 18 ...
Страница 47: ...47 ENGLISH 16 ELECTRICAL TEST Electrical safety test of earth ground and electrical insulation ...